IR vs. Raman microscopy – comparison, synergies and practical guidelines
- Abstract number
- 615
- Event
- European Microscopy Congress 2020
- DOI
- 10.22443/rms.emc2020.615
- Corresponding Email
- [email protected]
- Session
- LST.9 - Multiphoton imaging in highly scattering tissue
- Authors
- Alexander Kobald (2), Herbert Kobald (2), Claudia Mayrhofer (1), Harald Fitzek (2, 1)
- Affiliations
-
1. Graz Centre for Electron Microscopy (ZFE), Steyrergasse 17
2. Institute of Electron Microscopy and Nanoanalysis (FELMI), NAWI Graz, Graz University of Technology, Steyrergasse 17
- Keywords
Correlative Microscopy, FTIR, Raman
- Abstract text
We have compared Raman microscopy with the two main types of IR microscopy (Transmission, Attenuated Total Reflection) on five samples representing the typical areas of application, namely a packaging foil (polymers), a cigarette filter (particles), a paracetamol tablet (pharmaceuticals), a coated wood (biological materials) and a high-quality paper (organic-inorganic composite). Our goal is to give practical guidelines for the best choice of technique for any given application as well as highlight possible synergies from using more than one technique on the same sample.
IR and Raman microscopy are widely used techniques that are both based on vibrational spectroscopy. From a vibrational spectroscopic point of view, they yield complementary information due to their different selection rules, sensitivity to various compounds and limitations with regard to spectra range and background [1]. From a microscopic point of view (especially a practical one) the techniques have quite distinct advantages as well. The standard confocal Raman microscopy has a higher resolution, simpler sample preparation/handling and more versatility. Whereas the usual IR microscopy methods (ATR, Transmission) have higher measurement speeds and fewer background problems. We are discussing 3 questions in this work (for a given sample/ experimental situation): Which of the three microscopy techniques (confocal Raman, ATR-IR or Transmission-IR) is most practical? Which yields the most information? Is there an advantage of using more than one technique?
For the IR measurements a Bruker Tensor 27/Hyperion 3000 with a 64x64 focal-plane array detector using a x20 Ge-crystal objective (NA=0.4, ATR) and a x16 objective (NA=0.4, Transmission) was used. For the Raman measurements a LabRAM HR 800/Oylmpus BX 41 with various lasers and objectives was used. All data analysis was done with the respective instrument softwares and the spectral identification is based on experience and the KnowItAll® software/Bio-Rad database. The sample preparation was performed by embedding in epoxy resin and microtomy.
Let us look at the questions asked above using one of the five example (the coated wood). Figure 1 shows the Raman results and figure 2 the IR-ATR results. The IR-transmission results have been omitted for the sake of briefness, but show the distribution of lignin and cellulose on a much larger measurement area and with significantly worse resolution than both Raman and IR-ATR. The first and most important thing to mention about these results is that the Raman measurement took about 60 h due to the high fluorescence of the sample, whereas both IR measurements took only about 1 h each. So, if long measurement times are a problem Raman is a non-starter on this sample from a practical point of view. However, looking at the image quality Raman is clearly performing better than IR-ATR and is additionally identifying components of the coating, that were missed in the ATR measurement. Especially interesting is that the coloring dye (hostasol green) is found by Raman but not by IR, this is a general trend for coloring dyes, because they usually occur in very low concentration and Raman is especially sensitive to them due to Resonant Raman scattering [2]. Yet there are some areas with a high fluorescence (marked by yellow circles) that are difficult to identify completely with Raman microscopy. In these regions IR-ATR reveals grease and protein/human hair, presumably contaminations. This fits well into a more general trend, that it is usually easier to identify completely unkown components using IR-ATR, because there are more extensive IR database than Raman database and there are fewer background issues (like fluorescence in Raman).
To conclude the choice between Raman, IR-ATR and IR-transmission depends on many factors ranging from sample specific to theoretical and to get a good intuition about which method is best suited for your sample it is best to look at a range of examples and examine the comparison of the results and the practical aspects of the measurements in detail. This is what we aim to provide. In addition applying several techniques will in most cases give a more complete picture especially with regard to the sample composition, however the balance between better results and additional effort still has to be considered.
Figure 1: Results of the Raman mapping on the coated wood sample (right) and macroscopic/microscopic image of the sample (left top/bottom). Note that the regions marked with yellow circles show low signal to noise ratios and high fluorescence in the Raman measurement.
Figure 2: IR-ATR mappings of the same sample as shown in figure 1. Note that the sample is turned by 90 degrees. Left to right the integrated spectral regions are identified as grease (1756-1701 cm-1), protein/contamination with human hair (1703-1594 cm-1) and cellulose (1136-957 cm-1).
- References
[1] Larkin, P. (2017). Infrared and Raman spectroscopy: principles and spectral interpretation. Elsevier.
[2] Long, D. A.(2002). The Raman effect: a unified treatment of the theory of Raman scattering by molecules. John Wiley & Sons